Validation and determination of a reference interval for Canine HbA1c using an immunoturbidimetric assay

Abstract

Background: Hemoglobin A1c (HbA1c) provides a reliable measure of glycemic control over 2–3 months in human diabetes mellitus. In dogs, presence of HbA1c has been demonstrated, but there are no validated commercial assays. Objective: The purpose of the study was to validate a commercially available automated immunoturbidimetric assay for canine HbA1c and determine an RI in a hospital population. Methods: The specificity of the assay was assessed by inducing glycosylation in vitro using isolated canine hemoglobin, repeatability by measuring canine samples 5 times in succession, long term inter-assay imprecision by measuring supplied control materials, stability using samples stored at 4°C over 5 days and −20°C over 8 weeks, linearity by mixing samples of known HbA1c in differing proportions, and the effect of anticoagulants with paired samples. An RI was determined using EDTA-anticoagulated blood samples from 60 nondiabetic hospitalized animals of various ages and breeds. Hemoglobin A1c was also measured in 10 diabetic dogs. Results: The concentration of HbA1c increased proportionally with glucose concentration in vitro. For repeat measurements, the CV was 4.08% (range 1.16–6.10%). Samples were stable for 5 days at 4°C. The assay was linear within the assessed range. Heparin- and EDTA-anticoagulated blood provided comparable results. The RI for HbA1c was 9–18.5 mmol/mol. There was no apparent effect of age or breed on HbA1c. In diabetic dogs, HbA1c ranged from 14 to 48 mmol/mol. Conclusions: The assay provides a reliable method for canine HbA1c measurement with good analytic performance