An automated secondary standard for calibrating liquid flowmeters

Abstract

A secondary working standard of flow calibration has been developed to be used in place of a primary weight-time standard, and which can thereby effect a 75 percent reduction in calibration time while maintaining acceptable accuracies. The secondary standard uses six previously calibrated turbine-type flowmeters built into two manifold systems containing automatically switched flow valves. The pair of systems is capable of covering the flow range of 0.0004 to 19 l/s (0.007 to 300 gpm) with the uncertainty in volume flow rate not exceeding + or - 0.25 percent over the range of 0.06 to 19 l/s and not exceeding + or - 0.5 percent over the range 0.0004 to 0.06 1/s. Data reduction and plotting of results are by computer