research

Evaluation of volcano-style field ionization source and field emitting cathodes for mass spectrometry and applications

Abstract

A volcano-style field ionization source was tested with eight different gases: hydrogen, helium, ammonia, methane, argon, neon, water vapor, and hydrogen sulfide. For ammonia, hydrogen sulfide, and water, the ionization efficiency of the field ionization source was determined as a function of the electrical potential difference between the ionizer and its counterelectrode. The ionization efficiencies for the other gases were too low to be measured in the present apparatus. The operating characteristics of a field emission cathode, were studied, in the presence of the same eight gases at pressures up to 0.00001 torr. The presence of the gases caused little or no significant change in the electron emission from the cathodes. Results indicate that the field emission cathode has advantages over electrically heated cathodes as a source of an electron beam in spacecraft mass spectrometers