Initial dynamics of the EKG during an electrical defibrillation of the heart

Abstract

In tests on 11 mature dogs, immobilized by means of an automatic blocking and synchronization system, artefact free EKG were obtained, beginning 0.04-0.06 sec after passage of a defibrillating current. Different versions of the start of fibrillation were noted, in application of the defibrillating stimulus in the early phase of the cardiac cycle. A swinging phenomenon, increasing amplitude, of fibrillation was noted for 0.4-1.5 sec after delivery of a subthreshold stimulus. Conditions for a positive outcome of repeated defibrillation were found, and a relationship was noted between the configuration of the exciting process with respect to the lines of force of the defibrillating current and the defibrillation threshold. It was shown that the initial EKG dynamics after defibrillation is based on a gradual shift of the pacemaker from the myocardium of the ventricles to the sinus node, through phases of atrioventricular and atrial automatism

    Similar works