The Addition of EI/CI Capability to the Mattauch-herzog Spectrograph with EOID

Abstract

A modification was made to the Mattauch-Herzog Spectrograph with an electro-optical ion detector (EOID) previously designed and constructed, so that it would be capable of operating not only in the electron-impact (EI) mode of ionization, but also in the chemical ionization (CI) mode. This modification necessitated an effort in three specific design areas: (1) sample inlet; (2) ion source and analyzer regions; and (3) the pumping system. In addition, an appropriate electronics package had to be designed to control and operate the combined EI/CI source