research

Determination of physical and chemical states of lubricants in concentrated contacts, part 3

Abstract

Solid and liquid thin films were analyzed by infrared emission Fourier microspectrophotometry. The apparatus used is a commercial absorption instrument modified to an emission instrument, comprising a rotating polarizing device, a miniature blackbody temperature reference adjustable in temperature and radiant flux and a microscope lens with a high numerical aperture in the entrance system for increased sensitivity and resolution. Studies of lubricant behavior in a simulated ball bearing showed the alignment of the fluid molecules in the Hertzian area. Polyphenyl ether plus 1% 1,1,2-trichloroethane (TCE) required lower shear rates for the same degree of alignment than without TCE. The experiment was run with 440 C stainless steel balls coated with TiN, a chemically inert material. In both cases, the alignment was strongly influenced by the presence of TCE. The results showed (1) the dependence of alignment of fluid molecules on flow and not on adsorption at metallic surfaces, (2) phase separation between lubricant and additive under high pressure which results in two phase flow and (3) reduction in traction of torque transmitting (traction) fluids

    Similar works