research

A new calibration method for time delay standard and its application

Abstract

A method which is used to measure time delay accurately by using a Type 900-LB slotted line is described. The accuracy for calibrating time delay of a precision coaxial air line Type 900-L is about + or - (0.4 to 0.6)ps, and for coaxial cables with VSWR less than 1.5 and time delay t less than 50ns is about + or - (3 to 5)ps. Theoretical analysis and mathematical derivation of microwave networks in cascade are given. Methods to eliminate the errors which are caused by the discontinuities and the error analysis of the measuring system are presented. Skin effect analysis of the transient characteristic of coaxial transmission line are discussed in detail. Methods to eliminate the errors which result from using the calibrated time delay standard to calibrate time interval measurement instruments are presented. The estimation of errors and formulae for correction of those errors are described

    Similar works