Thematic Mapper image quality: Preliminary results

Abstract

Based on images analyzed so far, the band to band registration accuracy of the thematic mapper is very good. For bands within the same focal plane, the mean misregistrations are well within the specification, 0.2 pixels. For bands between the cooled and uncooled focal planes, there is a consistent mean misregistration of 0.5 pixels along-scan and 0.2-0.3 pixels across-scan. It exceeds the permitted 0.3 pixels for registration of bands between focal planes. If the mean misregistrations were removed by the data processing software, an analysis of the standard deviation of the misregistration indicates all band combinations would meet the registration specifications except for those including the thermal band. Analysis of the periodic noise in one image indicates a noise component in band 1 with a spatial frequency equivalent to 3.2 pixels in the along-scan direction