research

Comparison of modern icing cloud instruments

Abstract

Intercomparison tests with Particle Measuring Systems (PMS) were conducted. Cloud liquid water content (LWC) measurements were also taken with a Johnson and Williams (JW) hot-wire device and an icing rate device (Leigh IDS). Tests include varying cloud LWC (0.5 to 5 au gm), cloud median volume diameter (MVD) (15 to 26 microns), temperature (-29 to 20 C), and air speeds (50 to 285 mph). Comparisons were based upon evaluating probe estimates of cloud LWC and median volume diameter for given tunnel settings. Variations of plus or minus 10% and plus or minus 5% in LWC and MVD, respectively, were determined of spray clouds between test made at given tunnel settings (fixed LWC, MVD, and air speed) indicating cloud conditions were highly reproducible. Although LWC measurements from JW and Leigh devices were consistent with tunnel values, individual probe measurements either consistently over or underestimated tunnel values by factors ranging from about 0.2 to 2. Range amounted to a factor of 6 differences between LWC estimates of probes for given cloud conditions. For given cloud conditions, estimates of cloud MVD between probes were within plus or minus 3 microns and 93% of the test cases. Measurements overestimated tunnel values in the range between 10 to 20 microns. The need for improving currently used calibration procedures was indicated. Establishment of test facility (or facilities) such as an icing tunnel where instruments can be calibrated against known cloud standards would be a logical choice