Test and evaluation of the 2.4-micron photorefractor ocular screening system
- Publication date
- Publisher
Abstract
An improved 2.4-m photorefractor ocular screening system was tested and evaluated. The photorefractor system works on the principal of obtaining a colored photograph of both human eyes; and, by analysis of the retinal reflex images, certain ocular defects can be detected such a refractive error, strabismus, and lens obstructions. The 2.4-m photorefractory system uses a 35-mm camera with a telephoto lens and an electronic flash attachment. Retinal reflex images obtained from the new 2.4-m system are significantly improved over earlier systems in image quality. Other features were also improved, notably portability and reduction in mass. A total of 706 school age children were photorefracted, 211 learning disabled and 495 middle school students. The total students having abnormal retinal reflexes were 156 or 22 percent, and 133 or 85 percent of the abnormal had refractive error indicated. Ophthalmological examination was performed on 60 of these students and refractive error was verified in 57 or 95 percent of those examined. The new 2.4-m system has a NASA patent pending and is authorized by the FDA. It provides a reliable means of rapidly screening the eyes of children and young adults for vision problems. It is especially useful for infants and other non-communicative children who cannot be screened by the more conventional methods such as the familiar E chart