research

Wavelength error analysis in a multiple-beam Fizeau laser wavemeter having a linear diode array readout

Abstract

An estimate of the wavelength accuracy of a laser wavemeter is performed for a system consisting of a multiple-beam Fizeau interferometer and a linear photosensor array readout. The analysis consists of determining the fringe position errors which result when various noise sources are included in the fringe forming and detection process. Two methods of estimating the fringe centers are considered: (1) maximum pixel current location, and (2) average pixel location for two detectors with nearly equal output currents. Wavelength error results for these two methods are compared for some typical wavemeter parameters