Microchannel Plates for the UVCS and SUMER Instruments on the SOHO Satellite

Abstract

The microchannel plates for the detectors in the SUMER (Solar Ultraviolet Measurements of Emitted Radiation) and UVCS (Ultraviolet Coronograph Spectrometer) instruments aboard the Solar Orbiting Heliospheric Observatory (SOHO) mission to be launched in late 1995 are described. A low resistance Z stack of microchannel plates (MCP's) is employed in a detector format of 27 mm x 10 mm using a multilayer cross delay line anode (XDL) with 1024 x 360 digitized pixels. The MCP stacks provide gains of greater than 2 x 10(exp 7) with good pulse height distributions (as low as 25% FWHM) under uniform flood illumination. Background rates of approx. 0.6 event cm(exp -2) sec(exp -1) are obtained for this configuration. Local counting rates up to about 800 events/pixel/sec have been achieved with little drop of the MCP gain. MCP preconditioning results are discussed, showing that some MCP stacks fail to have gain decreases when subjected to a high flux UV scrub. Also, although the bare MCP quantum efficiencies are close to those expected (10%), we found that the long wavelength response of KBr photocathodes could be substantially enhanced by the MCP scrubbing process. Flat field images are characterized by a low level of MCP fixed pattern noise and are stable. Preliminary calibration results for the instruments are shown

    Similar works