slides

Future Development Trajectories for Imaging X-rays Spectrometers Based on Microcalorimeters

Abstract

Future development trajectories for imaging x-ray spectrometers based on microcalorimeters. Since their invention 30 years ago, the capability of X-ray microcalorimeters has increased steadily, with continual improvements in energy resolution, speed, and array size. Arrays of up to 1024 pixels have been produced, and resolution better than 1 eV at 1.5 keV has been achieved. These detectors can be optimized for the highest priority science, such as designing for the highest resolving power at low energies at the expense of dynamic range, or the greatest focal-plane coverage at the expense of speed. Three types of X-ray microcalorimeters presently dominate the field, each characterized by the thermometer technology. The first two types use temperature-sensitive resistors: semiconductors in the metal-insulator transition and superconductors operated in the superconducting-normal transition. The third type uses a magnetically coupled thermometer, and is at an earlier stage of development than the other two. The Soft X-ray Spectrometer (SXS) on Astro-H, expected to launch in 2015, will use an array of silicon thermistors with HgTe X-ray absorbers that will operate at 50 mK. Both the semiconductor and superconductor calorimeters have been implemented in small arrays. Kilopixel arrays of the superconducting calorimeters are being produced, and much larger arrays may require the non-dissipative advantage of magnetically coupled thermometers. I will project the development trajectories of these detectors and their read-out technologies and assess what their capabilities and limitations will be 10 - 20 years from now

    Similar works