Current 3D photoacoustic tomography (PAT) systems offer either high image
quality or high frame rates but are not able to deliver high spatial and
temporal resolution simultaneously, which limits their ability to image dynamic
processes in living tissue. A particular example is the planar Fabry-Perot (FP)
scanner, which yields high-resolution images but takes several minutes to
sequentially map the photoacoustic field on the sensor plane, point-by-point.
However, as the spatio-temporal complexity of many absorbing tissue structures
is rather low, the data recorded in such a conventional, regularly sampled
fashion is often highly redundant. We demonstrate that combining variational
image reconstruction methods using spatial sparsity constraints with the
development of novel PAT acquisition systems capable of sub-sampling the
acoustic wave field can dramatically increase the acquisition speed while
maintaining a good spatial resolution: First, we describe and model two general
spatial sub-sampling schemes. Then, we discuss how to implement them using the
FP scanner and demonstrate the potential of these novel compressed sensing PAT
devices through simulated data from a realistic numerical phantom and through
measured data from a dynamic experimental phantom as well as from in-vivo
experiments. Our results show that images with good spatial resolution and
contrast can be obtained from highly sub-sampled PAT data if variational image
reconstruction methods that describe the tissues structures with suitable
sparsity-constraints are used. In particular, we examine the use of total
variation regularization enhanced by Bregman iterations. These novel
reconstruction strategies offer new opportunities to dramatically increase the
acquisition speed of PAT scanners that employ point-by-point sequential
scanning as well as reducing the channel count of parallelized schemes that use
detector arrays.Comment: submitted to "Physics in Medicine and Biology