Iterative image reconstruction algorithms for optoacoustic tomography (OAT),
also known as photoacoustic tomography, have the ability to improve image
quality over analytic algorithms due to their ability to incorporate accurate
models of the imaging physics, instrument response, and measurement noise.
However, to date, there have been few reported attempts to employ advanced
iterative image reconstruction algorithms for improving image quality in
three-dimensional (3D) OAT. In this work, we implement and investigate two
iterative image reconstruction methods for use with a 3D OAT small animal
imager: namely, a penalized least-squares (PLS) method employing a quadratic
smoothness penalty and a PLS method employing a total variation norm penalty.
The reconstruction algorithms employ accurate models of the ultrasonic
transducer impulse responses. Experimental data sets are employed to compare
the performances of the iterative reconstruction algorithms to that of a 3D
filtered backprojection (FBP) algorithm. By use of quantitative measures of
image quality, we demonstrate that the iterative reconstruction algorithms can
mitigate image artifacts and preserve spatial resolution more effectively than
FBP algorithms. These features suggest that the use of advanced image
reconstruction algorithms can improve the effectiveness of 3D OAT while
reducing the amount of data required for biomedical applications