research

Measurement of rat insulin. Enzyme-linked immunosorbent assay with increased sensitivity, high accuracy, and greater practicability than established radioimmunoassay

Abstract

Total immunoreactive insulin (IRI) is conventionally determined by radioimmunoassays. IR! measurement in rats can be made more sensitive, accurate, and practical, as demonstrated by a new modified enzyme-linked immunosorbent assay (ELlSA). It is characterized by indirect binding of an anti-insulin antibody by an antiglobulin antibody and uses the principle of competitive saturation. In this ELlSA, IRI can be determined in a wide range of concentrations, corresponding to the standards. The standard curve ranges from 100 to 0.049 ng/mllRI (1 ng/ml - 23.4 JLU/ml - 172 pM rat insulin). The statistical analysis shows between- and within-assay coefficients of variation of :515%. Diabetes 37:321-26,198

    Similar works