The field of optical label free biosensors has become a topic of interest during past years, with devices based on the detection of angular or wavelength shift of optical modes [1]. Common parameters to characterize their performance are the Limit of Detection (LOD, defined as the minimum change of refractive index upon the sensing surface that the device is able to detect, and also BioLOD, which
represents the minimum amount of target analyte accurately resolved by the system; with units of concentration (common un its are p pm, ng/ml, or nM). LOD gives a first value to compare different biosensors, and is obtained both theoretically (using photonic calculation tools), and experimentally,covering the sensing area with fluids of different refractive indexes