In this paper we describe the development and first tests of a neutron
spectrometer designed for high flux environments, such as the ones found in
fast nuclear reactors. The spectrometer is based on the conversion of neutrons
impinging on 6Li into α and t whose total energy comprises the
initial neutron energy and the reaction Q-value. The 6LiF layer is
sandwiched between two CVD diamond detectors, which measure the two reaction
products in coincidence. The spectrometer was calibrated at two neutron
energies in well known thermal and 3 MeV neutron fluxes. The measured neutron
detection efficiency varies from 4.2×10−4 to 3.5×10−8 for
thermal and 3 MeV neutrons, respectively. These values are in agreement with
Geant4 simulations and close to simple estimates based on the knowledge of the
6Li(n,α)t cross section. The energy resolution of the spectrometer
was found to be better than 100 keV when using 5 m cables between the detector
and the preamplifiers.Comment: submitted to NI