IQMDose3D: a software tool for reconstructing the dose in patient using patient planning CT images and the signals measured by IQM detector

Abstract

The integral quality monitor (IQM) system compares the signal measured with a large volume chamber mounted to the linear accelerator's head to the signal calculated using the patient DICOM RT plan for patient-specific quality assurance (PSQA). A method was developed to reconstruct the dose in patients using the signal measured by IQM chamber and patient planning CT images. A software tool named IQMDose3D was implemented to automate this procedure and integrated into the IQM-based PSQA workflow. IQMDose3D enables the physicists to evaluate PSQA by focusing on the clinical perspective by comparing the delivered plan to the approved clinical plan in terms of the clinical goals, dose-volume histogram (DVH) in addition to the three-dimensional (3D) gamma map and gamma pass rate.Comment: Accepted by ICCR 2024 conferenc