A Leadless Ventricular Pacemaker Providing Atrioventricular Synchronous Pacing in the Real-World Setting: 12-Month Results from the Micra AV Post-Approval Registry.

Abstract

BACKGROUND Advances in leadless pacemaker technology have enabled accelerometer-based atrioventricular (AV) synchronous pacing by sensing atrial mechanical contraction. OBJECTIVES To report performance of the Micra AV leadless pacemaker from the worldwide Micra AV post-approval registry (PAR) through 12-months. METHODS The Micra AV PAR is a prospective single-arm observational registry designed to assess safety and effectiveness of Micra AV in a real-world setting. For the present interim analysis, major complications and system revisions through 12-months were summarized and compared to a historical cohort of 2,667 transvenous dual-chamber pacing patients. RESULTS The device was successfully implanted in 796 of 801 patients (99.4%) at 97 centers in 19 countries. Micra AV patients were older (74.1 vs. 71.1 years, P90%. CONCLUSIONS The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple co-morbidities, with a significantly lower rate of complications and system revisions through 12-months compared to a historical cohort of patients with transvenous dual-chamber pacemakers