Macular thickness measurements in normal eyes using spectral domain optical coherence tomography

Abstract

Knowledge of the macular thickness in a normal population is important for the evaluation of pathological macular change. The purpose of this study was to define and measure macular thickness in normal eyes using spectral domain optical coherence tomography (OCT). Fifty eyes from 50 normal subjects (29 men and 21 women, aged 22 to 68 years) were scanned with a prototype Cirrus HD-OCT system (5 microm axial resolution) (Carl Zeiss Meditec, Inc.). The proprietary Cirrus segmentation algorithm was used to produce retinal thickness maps, which were then averaged over 9 regions defined by a circular target centered at the true fovea location. The macular thickness of 13 subjects scanned with both HD-OCT and StratusOCT were compared. After centering the fovea, the mean and standard deviation values for retinal thickness measurements were calculated point wise and averaged on standard regions. For patients scanned with both systems, the thickness measurements from HD-OCT were approximately 50 microm larger than those from StratusOCT. The difference between the two measurements decreased somewhat with eccentricity. Using HD-OCT, it is possible to acquire retinal data sets containing an unprecedented number of data points. Furthermore, it is possible to use OCT fundus images to evaluate the scan quality and to center the measurement at the fovea. These advantages, together with good automated segmentation, can produce more accurate retinal thickness measurements. Incorporation of the photoreceptor layer in the measurements is anatomically meaningful and may be significant in evaluating various retinal pathologies and visual acuity outcomes

    Similar works