Can a Pb/SCIFI calorimeter survive the SSC

Abstract

A scintillating fiber based electromagnetic calorimeter module built from radiation-hard materials has been tested in a beam capable of delivering both low and high currents of monoenergetic electrons. Energy resolution and light output measurements were made following high-dose exposures. The procedure was repeated until the resolution of the detector decreased from an initial value of 6.9%{radical}E to 14.0%{radical}E and the pulse height dropped by a factor of 11. After four weeks, the detector was retested. Partial recovery was observed in the light output which returned to approximately 52% of its original value. The resolution recovered to a value of 8.8%{radical}E. The tests are described. 9 refs., 4 figs

    Similar works