Monte Carlo characterization of the AAPM Task Group 43 Dosimetry Parameters for the M-15 high dose rate Iridium-192 brachytherapy source

Abstract

The Source Production & Equipment Co. (SPEC) Model M-15 Iridium Brachytherapy is intended for use as a temporary High Dose Rate (HDR) Brachytherapy source. The clinical dosimetry parameters of this source was calculated using the MCNP Monte Carlo analysis program (MCNP6; Los Alamos National Laboratory) in accordance to the updated AAPM Task Group Report No.43 (TG43U1). All TG43U1 dosimetry parameters and their respective statistical uncertainty are presented in the reports

    Similar works

    Full text

    thumbnail-image

    Available Versions