Calculations for Calibration of a Mass Spectrometer

Abstract

A computer program performs calculations to calibrate a quadrupole mass spectrometer in an instrumentation system for identifying trace amounts of organic chemicals in air. In the operation of the mass spectrometer, the mass-to-charge ratio (m/z) of ions being counted at a given instant of time is a function of the instantaneous value of a repeating ramp voltage waveform applied to electrodes. The count rate as a function of time can be converted to an m/z spectrum (equivalent to a mass spectrum for singly charged ions), provided that a calibration of m/z is available. The present computer program can perform the calibration in either or both of two ways: (1) Following a data-based approach, it can utilize the count-rate peaks and the times thereof measured when fed with air containing known organic compounds. (2) It can utilize a theoretical proportionality between the instantaneous m/z and the instantaneous value of an oscillating applied voltage. The program can also estimate the error of the calibration performed by the data-based approach. If calibrations are performed in both ways, then the results can be compared to obtain further estimates of errors