In order to better build the Neutral Beam Injector with Negative Ion Source
(NNBI), the pre-research on key technologies has been carried out for the
Comprehensive Research Facility for Fusion Technology (CRAFT). Cesium seeding
into negative-ion sources is a prerequisite to obtain the required negative
hydrogen ion. The performance of ion source largely depends on the cesium
conditions in the source. It is very necessary to quantitatively measure the
amount of cesium in the source during the plasma on and off periods (vacuum
stage). This article uses the absorption peak of cesium atoms near 852.1nm to
build a cesium atom concentration detection system based on Tunable Diode Laser
Absorption Spectroscopy (TDLAS) technology. The test experiment based on the
cesium cell is carried out, obtained the variation curve of cesium
concentration at different temperatures. The experimental results indicate
that: the system detection range is within 5*10E6-2.5*10E7 pieces/cm3 and the
system resolution better than 1*10E6 pieces/cm3.Comment: 8 pages,7 figures, the 20th International Symposium on Laser-Aided
Plasma Diagnostic