Aim: To assess the medium term outcomes of Acrysof® toric intraocular lens implantation
in 54 patients (54 eyes).
Methods: Uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), preoperative
astigmatism, residual postoperative astigmatism, and global average and model-specific
intraocular lens (IOL) rotation grade were analyzed.
Results: At 2-months of follow-up,the mean UCVA was 0.83 (SD: 0.14) Snellen scale, with
73.9% of the patients ≥0.8, and 32.6% with 1.0. Mean BCVA achieved was 0.94 (SD: 0.10).
Mean preoperatory astigmatism was –2.25 diopters (D) (SD: 0.78), and mean postoperative
astigmatism was -0.32 D (SD: 0.56), with significant differences between both groups
(p<0.001). Model-specific mean residual astigmatism was –0.1 D for T3, –0.27 D for T4 and
–0.43 D for T5, without significant differences between the three models (p=0.483). Mean
IOL-axis rotation grade was 3.87±3.25 degrees, with 91.6% of implanted lens within 10° of
predicted axis.
Discussion: T3, T4 and T5 Acrysof® Toric intraocular lenses can correct preoperative
astigmatism with a high success rate in terms of UCVA, and residual postoperative
astigmatism, with minimum IOL-rotation grade at 2 months follow-up period