Quantitative portable gamma spectroscopy sample analysis for non-standard sample geometries

Abstract

Utilizing a portable spectroscopy system, a quantitative method for analysis of samples containing a mixture of fission and activation products in nonstandard geometries was developed. The method can be used with various sample and shielding configurations where analysis on a laboratory based gamma spectroscopy system is impractical. The portable gamma spectroscopy method involves calibration of the detector and modeling of the sample and shielding to identify and quantify the radionuclides present in the sample. The method utilizes the intrinsic efficiency of the detector and the unattenuated gamma fluence rate at the detector surface per unit activity from the sample to calculate the nuclide activity and Minimum Detectable Activity (MDA). For a complex geometry, a computer code written for shielding applications (MICROSHIELD) is utilized to determine the unattenuated gamma fluence rate per unit activity at the detector surface. Lastly, the method is only applicable to nuclides which emit gamma rays and cannot be used for pure beta emitters. In addition, if sample self absorption and shielding is significant, the attenuation will result in high MDA`s for nuclides which solely emit low energy gamma rays. The following presents the analysis technique and presents verification results demonstrating the accuracy of the method

    Similar works