Macrostrain measurement using radial collimators at LANSCE

Abstract

A series of `short` radial collimators have been implemented in the 90{degrees} scattering geometries on the neutron powder diffractometer at Los Alamos. The capability to perform macrostrain measurements has been improved by the commensurate ability to rapidly select a sampling volume appropriate to the specimen. The compact design of the collimators was dictated by the need to fit them in a cylindrical vacuum chamber as well as providing space in which to manipulate a specimen in three dimensions. Collimators of different vane lengths were fabricated to give 4 different resolutions for which 2/3 of the diffracted intensity comes form distances of 0.75, 1. 25, 2.5, and 4.0 mm along the incident beam. Qualifying scans and a demonstration of a cracked ring, containing a steep stress gradient, are included