The standard point-of-care Hemochron Jr. ACT+ test in monitoring heparin administration for cardiopulmonary bypass in severe factor XII deficiency

Abstract

Abstract Coagulation factor XII (FXII) is a plasma serine protease that belongs to the contact activation complex responsible for initiating the intrinsic coagulation pathway. FXII deficiency is a rare congenital disorder that is not associated with an increased tendency for bleeding. However, as contact activation is impaired in FXII deficiency, both the celite- and kaolin-initiated activated clotting time (ACT) measurements are prolonged markedly, which poses a challenge for anticoagulation monitoring in patients undergoing cardiac surgery. The authors successfully have used the standard Hemochron Jr. ACT+ test, which is activated by silica and phospholipid in addition to kaolin, to monitor anticoagulation for cardiopulmonary bypass in two patients with severe FXII deficiency. The ACT+ test showed low baseline values, increased adequately in response to heparin, and decreased to baseline after protamine. Importantly, there was no abnormal intra- or postoperative bleeding nor any thrombotic complications. Furthermore, in vitro dose-response ACT+ testing of FXII-deficient blood with increasing heparin concentrations supports the use of ACT+ in FXII deficiency

    Similar works

    Full text

    thumbnail-image