SIMPLE HPLC-UV METHOD FOR DETERMINATION OF METFORMIN IN HUMAN PLASMA AND ERYTHROCYTES APPLICATION TO THERAPEUTIC DRUG MONITORING

Abstract

Objective: The aim of this study was to develop a simple, rapid, efficient, cost effective and reproducible, stability indicating reverse phase high performance liquid chromatography method (RP-HPLC) for dosage of metformin in human plasma and erythrocytes. Methods: In this method, the plasma or erythrocyte proteins were precipitated using Perchloric acid: acetonitrile (50 % v/v) mixture and the supernatant liquid were injected into the HPLC system. The separation was achieved with a symmetry C8 column with the mobile phase containing 10 % water and 90 % sodium dihydrogen phosphate buffer (5.8 mM), the pH was adjusted to 3.8 with Phosphoric acid. The temperature was elevated to 25 °C. The detection was done by a UV detector at 232 nm. Results: The retention time was observed at around 4.412 min for metformin and 6.022 for lansoprazole an internal standard (IS). The response was linear over a range of 2-32µg ml-1, the coefficient of determination (r²) was found to be (r² =0. 9988). The lowest limit of quantification and detection was 0.1 µg/ml and 0.3 µg/ml respectively. No endogenous substances were found to interfere with the peaks of the drug. The intra-day and inter-day coefficient of variations was 2.1 % or less for all the selected concentrations. The relative errors at all the studied concentrations were 3.5 % or less. Conclusion: The HPLC method described in this article was simple, selective, reproducible, linear, and precise, it can be applied for therapeutic drug monitoring of metformin in human plasma and erythrocytes

    Similar works