A homogeneous HDL-c assay (HDL-H), which uses polyethylene glycol-modified
enzymes and sulfated alpha-cyclodextrin, was assessed for precision,
accuracy, and cholesterol and triglyceride interference. In addition, its
analytical performance was compared with that of a phosphotungstic acid
(PTA)/MgCl2 precipitation method (HDL-P). Within-run CVs were < or =
1.87%; total CVs were < or = 3.08%. Accuracy was evaluated in fresh
normotriglyceridemic sera using the Designated Comparison Method (HDL-H =
1.037 Designated Comparison Method + 4 mg/L; n = 63) and in moderately
hypertriglyceridemic sera by using the Reference Method (HDL-H = 1.068
Reference Method - 17 mg/L; n = 41). Mean biases were 4.5% and 2.2%,
respectively. In hypertriglyceridemic sera (n = 85), HDL-H concentrations
were increasingly positively biased with increasing triglyceride
concentrations. The method comparison between HDL-H and HDL-P yielded the
following equation: HDL-H = 1.037 HDL-P + 15 mg/L; n = 478. We conclude
that HDL-H amply meets the 1998 NCEP recommendations for total error; its
precision is superior compared with that of HDL-P, and its average bias
remains below +/-5% as long as triglyceride concentrations are < or = 10
g/L and in case of moderate hypercholesterolemia