Iterative RF pulse design for multidimensional, small-tip-angle selective excitation

Abstract

The excitation k -space perspective on small-tip-angle selective excitation has facilitated RF pulse designs in a range of MR applications. In this paper, k -space-based design of multidimensional RF pulses is formulated as a quadratic optimization problem, and solved efficiently by the iterative conjugate-gradient (CG) algorithm. Compared to conventional design approaches, such as the conjugate-phase (CP) method, the new design approach is beneficial in several regards. It generally produces more accurate excitation patterns. The improvement is particularly significant when k -space is undersampled, and it can potentially shorten pulse lengths. A prominent improvement in accuracy is also observed when large off-resonance gradients are present. A further boost in excitation accuracy can be accomplished in regions of interest (ROIs) if they are specified together with “don't-care” regions. The density compensation function (DCF) is no longer required. In addition, regularization techniques allow control over integrated and peak pulse power. Magn Reson Med, 2005. © 2005 Wiley-Liss, Inc.Peer Reviewedhttp://deepblue.lib.umich.edu/bitstream/2027.42/48766/1/20631_ftp.pd