Scintillation light from gamma ray irradiation in liquid xenon is detected by
two Hamamatsu R9288 photomultiplier tubes (PMTs) immersed in the liquid. UV
light reflector material, PTFE, is used to optimize the light collection
efficiency. The detector gives a high light yield of 6 photoelectron per keV
(pe/keV), which allows efficient detection of the 122 keV gamma-ray line from
Co-57, with a measured energy resolution of (8.8+/-0.6)% (sigma). The best
achievable energy resolution, by removing the instrumental fluctuations, from
liquid xenon scintillation light is estimated to be around 6-8% (sigma) for
gamma-ray with energy between 662 keV and 122 keV