A liquid isooctane (C8H18) filled ionization linear array for
radiotherapy quality assurance has been designed, built and tested. The
detector consists of 128 pixels, each of them with an area of 1.7 mm ×
1.7 mm and a gap of 0.5 mm. The small pixel size makes the detector ideal for
high gradient beam profiles like those present in Intensity Modulated Radiation
Therapy (IMRT) and radiosurgery. As read-out electronics we use the X-Ray Data
Acquisition System (XDAS) with the Xchip developed by the CCLRC.
Studies concerning the collection efficiency dependence on the polarization
voltage and on the dose rate have been made in order to optimize the device
operation.
In the first tests we have studied dose rate and energy dependences, and
signal reproducibility. Dose rate dependence was found lower than 2.5 % up to 5
Gy min−1, and energy dependence lower than 2.1 % up to 20 cm depth in
solid water. Output factors and penumbras for several rectangular fields have
been measured with the linear array and were compared with the results obtained
with a 0.125 cm3 air ionization chamber and radiographic film,
respectively. Finally, we have acquired profiles for an IMRT field and for a
virtual wedge. These profiles have also been compared with radiographic film
measurements. All the comparisons show a good correspondence. Signal
reproducibility was within a 2% during the test period (around three months).
The device has proved its capability to verify on-line therapy beams with
good spatial resolution and signal to noise ratio.Comment: 16 pages, 12 figures Submitted to Phys. Med. Bio