We have exposed seven full length production quality crystals of the
electromagnetic calorimeter (ECAL) of the CMS detector to a 20 GeV/c proton
beam at the CERN PS accelerator. The exposure was done at fluxes of 10**12
p/cm**2/h and 10**13 p/cm**2/h and integral fluences of 10**12 p/cm**2 and
10**13 p/cm**2 were reached at both rates. The light transmission of the
crystals was measured after irradiation and suitable cooling time for induced
radioactivity to decrease to a safe level. First results of these measurements
are shown. The possible damage mechanisms are discussed and simulations based
on one possible model are presented. The implications for long-term operation
of CMS are discussed and it is shown that in the whole barrel and at least most
of the ECAL endcap hadron damage alone - even if cumulative - should not cause
the crystals to fail the CMS specification of an induced absorption coefficient
muIND < 1.5 /m during the first 10 years of LHC operation.Comment: 5 pages, to be published in Proc. ICATPP Conference on Astroparticle,
Particle, Space Physics, Detectors and Medical Physics Applications (Como,
Italy, 6 to 10 October 2003