We have used a mesh experiment in order to measure the sub-pixel structure of
the EPIC MOS CCDs on-board the XMM/NEWTON satellite. The EPIC MOS CCDs have 40
μm-square pixels which have an open electrode structure in order to improve
the detection efficiency for low-energy X-rays. We obtained restored pixel
images for various X-ray event grades (e.g. split-pixel events, single pixel
events, etc.) at various X-ray energies.
We confirmed that the open electrode structure results in a distorted
horizontal pixel boundary. The open electrode region generates both single
pixel events and vertically split events, but no horizontally split events.
Because the single pixel events usually show the best energy resolution, we
discuss a method of increasing the fraction of single pixel events from the
open electrode region. Furthermore, we have directly measured the thickness of
the electrodes and dead-layers by comparing spectra from the open electrode
region with those from the other regions: electrodes, electrode finger and
channel stop. We can say that EPIC MOS CCDs are more radiation hard than
front-illumination chips of ACIS on-board Chandra X-ray Observatory because of
their extra absorption thickness above the charge transfer channel. We
calcurated the mean pixel response and found that our estimation has a good
agreement with that of the ground calibration of EPIC MOS CCD.Comment: 20pages including 2 tables, 10 figures,Accepted for publication in :
Nuclear Instruments and Methods in Physics Research