Univerzitet u Beogradu - Institut za nuklearne nauke Vinča, Beograd
Abstract
The effect of the density of environmental samples on the counting efficiency of Ge detectors used in gamma-spectrometry was studied. The dependence εff(ρ) was determined for two HPGe detectors (relative efficiencies 18% and 20%) using five radioactive standard reference materials (silicone resins, epoxy resin, milk powder, soil) with different matrix densities (0.45-1.22 g/cm3) in Marinelli beakers (V = 500 cm3). The dependence of efficiency vs. density was found to be linear and the regression parameters for energies in the range of 60-2000 keV were determined, too. The effect of variation in density on the counting efficiency of Ge detectors is dominant in the range of lower energies (60-600 keV) and de creases with energies in the higher energy range.Ispitivan je uticaj gustine uzoraka iz životne sredine na efikasnost brojanja germanijumskih detektora u spektrometriji gama zračenja. Zavisnost εff(ρ) određena je za dva HPGe detektora (relativnih efikasnosti 18% i 20%) za različite radioaktivne standardne referentne materijale (silikonske smole epoksi smolu, mleko u prahu, zemlju) različitih gustina (0,45-1,22 g/cm3) u Marineli posudama zapremine V = 500 cm3. Utvrđena je linearna zavisnost efikasnosti brojanja od gustine uzoraka pri čemu su određeni parametri linearne regresije za niz energija iz intervala od 60-2000 keV. Efekt promene gustina na efikasnost brojanja germanijumskih detektora dominantan je u intervalu nižih energija (60-600 keV) i smanjuje se sa porastom energije u intervalu viših energija