It is essential for medical diagnoses to analyze Electrocardiogram (ECG signal). The core of this analysis is to detect the QRS complex. A modified approach is suggested in this work for QRS detection of ECG signals using existing database of arrhythmias. The proposed approach starts withΒ Β the same steps of previous approaches by filtering the ECG. The filtered signal is then fed to a differentiator to enhance the signal. The modified adaptive threshold method which is suggested in this work, is used to detect QRS complex. This method uses a new approach for adapting threshold level, which is based on statistical analysis of the signal. Forty-eight records from an existing arrhythmia database have been tested using the modified method. The result of the proposed method shows the high performance metrics with sensitivity of 99.62% and a positive predictivity of 99.88% for QRS complex detection