Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers

Abstract

Background. Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting. We investigated whether the prothrombin time international normalized ratio (INR) results by CoaguChek XS Plus (Roche Diagnostics GmbH, Mannheim, Germany) can be used reliably without being con�rmed with the INR results by STA-R system (Diagnostica Stago S.A.S, Asnières sur Seine, France). Methods. A total of 118 INR measurements by CoaguChek XS Plus and STA-R were compared using Passing/Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurements was further assessed in relation to dosing decision. Results. e correlation of INR measurements between CoaguChek XS Plus and STA-R was excellent (correlation coefficient = 0.964). e mean difference tended to increase as INR results increased and was 0.25 INR in the therapeutic range (2.0-3.0 INR). e overall agreement was fair to good (kappa = 0.679), and 21/118 (17.8%) INR measurements showed a difference in dosing decision. Conclusion. e positive bias of CoaguChek XS Plus may be obvious even in the therapeutic INR range, and dosing decision based on the CoaguChek XS Plus INR results would be different from that based on the STA-R results. e INR measurements by POCT coagulometers still need to be con�rmed with the laboratory INR measurements