Differences in crestal bone-to-implant contact following an under-drilling compared to an over-drilling protocol. A study in the rabbit tibia

Abstract

The objective of this study is to compare bone-to-implant contact (BIC) between implants inserted at high torque due to under-drilling of the crestal bone to those inserted at low torque due to over-drilling of the crestal bone. Forty implants with diameters of 3.75 mm (group A) or 3.55 mm (group B) were inserted in the proximal tibiae of NZW rabbits in two separate surgeries on day 0 or 21. Osteotomy of the crestal bone was finalized with a 3.65-mm drill. In group A, implants were inserted at torque ae35 Ncm (under-drilling) and in group B with torque 0.05). No differences in t-BIC were noted at 3 weeks and at 6 weeks between the groups. Insertion of implants with an over-drilling protocol of the crestal aspect of the osteotomy resulted in increased short-term crestal bone-to-implant contact. Insertion of implants with a high torque following an under-drilling protocol, commonly used for immediate loading, may reduce crestal bone-to-implant contact at early healing stages