    Spring distraction system for dynamic growth guidance of early onset scoliosis: two-year prospective follow-up of 24 patients

    Get PDF
    BACKGROUND: Current surgical treatment options for early onset scoliosis (EOS), with distraction- or growth-guidance implants, show limited growth and high complication rates during follow-up. We developed a novel implant concept, which uses compressed helical springs positioned around the rods of a growth-guidance construct. This spring distraction system (SDS) provides continuous corrective force to stimulate spinal growth, can be easily contoured, and can be used with all standard spinal instrumentation systems. PURPOSE: To assess curve correction and -maintenance, spinal growth, complication rate, and health-related quality of life following SDS treatment. STUDY DESIGN: Prospective cohort study. PATIENT SAMPLE: All skeletally immature EOS patients with an indication for growth-friendly surgery and without bone- or soft tissue weakness were eligible to receive SDS. For this study, all included patients with at least 2-year follow-up were analyzed. OUTCOME MEASURES: Coronal Cobb angle, sagittal parameters, T1-T12, T1-S1, and instrumented (ie, bridged segment) spinal height and freehand length, complications and re-operations, and the 24-Item Early Onset Scoliosis Questionnaires (EOSQ-24) score. METHODS: All primary- and conversion patients (conversion from failed other systems) with SDS and ≥2 years follow-up were included. Radiographic parameters were compared preoperatively, postoperatively and at latest follow-up. Spinal length increase was expressed as mm/year. RESULTS: Twenty-four skeletally immature EOS patients (18 primary and 6 conversion cases) were included. There were five idiopathic, seven congenital, three syndromic, and nine neuromuscular EOS patients. Mean age at implantation was 9.1 years (primary: 8.4; conversion: 11.2). Major curve improved from 60.3° to 35.3°, and was maintained at 40.6° at latest follow-up. Mean spring length increase during follow-up was 10.4 mm/year. T1-S1 height increased 9.9mm/year and the instrumented segment height showed a mean increase of 0.7 mm/segment/year. EOSQ-24 scores dropped after surgery from 75.6 to 67.4 but recovered to 75.0 at latest follow-up. In total, 17 reoperations were performed. Ten reoperations were performed to treat 9 implant-related complications. In addition, 7 patients showed spinal growth that exceeded expected growth velocity; their springs were retensioned during a small reoperation. CONCLUSION: The 2-year follow-up results from this prospective cohort study indicate that the concept of spring distraction may be feasible as an alternative to current growing spine solutions. Curve correction and growth could be maintained satisfactory without the need for repetitive lengthening procedures. However, as in all growth-friendly implants, complications and reoperations could not be prevented, which emphasizes the need for further improvement

    The potential of spring distraction to dynamically correct complex spinal deformities in the growing child

    Get PDF
    Purpose: Current treatment of progressive early onset scoliosis involves growth-friendly instrumentation if conservative treatment fails. These implants guide growth by passive sliding or repeated lengthenings. None of these techniques provide dynamic correction after implantation. We developed the spring distraction system (SDS), by using one or multiple compressed springs positioned around a standard sliding rod, to provide active continuous distraction of the spine to stimulate growth and further correction. The purpose of this study was to determine feasibility and proof of concept of the SDS. Methods: We developed a versatile, dynamic spring distraction system for patients who would benefit from active continuous distraction. This prospective case series evaluates four patients with exceptional and progressive congenital spine deformities. Results: Four patients had a mean age of 6.8 years at surgery with a mean follow-up of 36 months (range 25–45). The mean progressive thoracic lordosis, which was the reason for initiating surgical treatment in two patients, changed from 32° lordosis preoperatively to 1° kyphosis post-operatively. During follow-up, this further improved to 32° thoracic kyphosis. In the two other patients, with cervicothorcacic scoliosis, the main coronal curve improved from 79° pre-operatively to 56° post-operatively and further improved to 42°. The mean T1-S1 spine growth during follow-up for all patients was 1.3 cm/year. There was one reoperation because of skin problems and no device-failures. Conclusion: These early results show the feasibility and the proof of concept of spring-based distraction as a dynamic growth-enhancing system with the potential of further correction of the deformity after implantation

    Induction of a representative idiopathic-like scoliosis in a porcine model using a multidirectional dynamic spring-based system

    Get PDF
    BACKGROUND CONTEXT: Scoliosis is a 3D deformity of the spine in which vertebral rotation plays an important role. However, no treatment strategy currently exists that primarily applies a continuous rotational moment over a long period of time to the spine, while preserving its mobility. We developed a dynamic, torsional device that can be inserted with standard posterior instrumentation. The feasibility of this implant to rotate the spine and preserve motion was tested in growing mini-pigs. PURPOSE: To test the quality and feasibility of the torsional device to induce the typical axial rotation of scoliosis while maintaining growth and mobility of the spine. STUDY DESIGN: Preclinical animal study with 14 male, 7 month old Gottingen mini-pigs. Comparison of two scoliosis induction methods, with and without the torsional device, with respect to 3D deformity and maintenance of the scoliosis after removal of the implants. METHODS: Fourteen mini-pigs received either a unilateral tether-only (n=6) or a tether combined with a contralateral torsional device (n=8). X-rays and CT-scans were made post-operative, at 8 weeks and at 12 weeks. Flexibility of the spine was assessed at 12 weeks. In 3 mini-pigs per condition, the implants were removed and the animals were followed until no further correction was expected. RESULTS: At 12 weeks the tether-only group yielded a coronal Cobb angle of 16.8±3.3°For the tether combined with the torsional device this was 22.0±4.0°. The most prominent difference at 12 weeks was the axial rotation with 3.6±2.8° for the tether-only group compared to 18.1±4.6° for the tether-torsion group. Spinal growth and flexibility remained normal and comparable for both groups. After removal of the devices, the induced scoliosis reduced by 41% in both groups. There were no adverse tissue reactions, implant complications or infections. CONCLUSION: The present study indicates the ability of the torsional device combined with a tether to induce a flexible idiopathic-like scoliosis in mini-pigs. The torsional device was necessary to induce the typical axial rotation found in human scoliosis. Clinical significance: The investigated torsional device could induce apical rotation in a flexible and growing spine. Whether this may be used to reduce a scoliotic deformity remains to be investigated

    Intrawound Treatment for Prevention of Surgical Site Infections in Instrumented Spinal Surgery : A Systematic Comparative Effectiveness Review and Meta-Analysis

    No full text
    Study Design: Systematic review and meta-analysis. Objectives: To determine the efficacy of intrawound treatments in reducing deep surgical site infections (SSIs) in instrumented spinal surgery. Methods: The electronic databases MEDLINE, EMBASE, and Cochrane were systematically searched for intrawound treatments for the prevention of SSIs in clean instrumented spine surgery. Both randomized controlled trials and comparative cohort studies were included. The results of included studies were pooled for meta-analysis. Results: After full text- and reference screening, 20 articles were included. There were 2 randomized controlled trials and 18 observational studies. Sixteen studies investigated the use of intrawound antibiotics, and 4 studies investigated the use of intrawound antiseptics. The relative risk of deep SSI for any treatment was 0.26 (95% confidence interval [CI] 0.16-0.44, P < .0001), a significant reduction compared with controls receiving no treatment. For patients treated with local antibiotics the relative risk was 0.29 (95% CI 0.17-0.51, P < .0001), and patients treated with local antiseptics had a relative risk of 0.14 (95% CI 0.05-0.44, P = .0006). Conclusions: Both the use of antibiotic and antiseptic intrawound prophylactics was associated with a significant 3 to 7 times reduction of deep SSIs in instrumented spine surgery. No adverse events were reported in the included studies

    The Spring Distraction System for Growth-Friendly Surgical Treatment of Early Onset Scoliosis: A Preliminary Report on Clinical Results and Safety after Design Iterations in a Prospective Clinical Trial

    Get PDF
    Background: The Spring Distraction System (SDS) is a dynamic growth-friendly implant to treat early onset scoliosis (EOS). Previous SDS studies showed promising results in terms of curve correction and complication profile. Nevertheless, complications did occur, which led to modifications in the implant design. The main iterations were a larger rod diameter and a more sagittal stable sliding mechanism. The purpose of this study was to investigate the performance of these iterations. Methods: All patients treated with the modified SDS and &gt;1 year follow-up were included. Radiographic outcomes, severe adverse events (SAEs), unplanned returns to the operating room (UPRORs) and health-related quality of life (HRQoL) were investigated. Results: Seventeen EOS patients (three congenital, four idiopathic, nine neuromuscular, one syndromic) were included. Mean age at surgery was 9.5 &plusmn; 2.5 years. Similar to the first generation SDS, about 50% initial correction was achieved and maintained, and spinal growth was near physiological. Most importantly, SAEs and UPRORs were diminished and favorable with 0.10/patient/year. In addition, HRQoL increased during the first year postoperatively, indicating the implant was well accepted. Conclusion: These preliminary results indicate that the iterations of the SDS are effective in terms of reducing SAEs and UPRORs and increasing HRQoL in patients with EOS

    Disc and Vertebral Body Morphology From Birth to Adulthood

    No full text
    Study Design.Cross-sectional.Objective.The aim of this study was to describe the morphology of intervertebral discs and vertebral bodies during growth in asymptomatic children and adolescents.Summary of Background Data.Earlier studies demonstrated that spinal growth occurs predominantly in vertebral bodies. This axiom introduced a vertebral-body-focus for unravelling etiological questions and achieve growth-modulation in young spinal deformity patients. Recent studies show the importance of the intervertebral discs in the early phases and possible etiology of pediatric spinal deformities. There is presently a paucity of 3D morphometric data of spinal elements during growth.Methods.A database of 298 patients aged 0 to 21 that have received a computed tomography scan for indications not related to the spine was analyzed. Custom made software was used to semi-automatically measure intervertebral disc and vertebral body morphology, corrected for orientation in all 3 planes.Results.Vertebral body height increased from birth up to adulthood, from 4-to-14 mm in the cervical, 6 to 20 mm in the thoracic, and 9 to 28 mm in the lumbar spine. This increase was 0.70 mm/year in males, more pronounced than females with 0.62 mm/year (P = 0.001). Lumbar discs increased throughout growth from 4.4 to 9.0 mm, whereas thoracic discs only increased from 3.5 to 4.9 mm at age 4 and remained stable afterwards, similarly for cervical discs. The disc transverse surface area increased greatly and consistently throughout growth. Disc slenderness was stable in the lumbar spine during growth, but decreased in the thoracic and cervical spine. Overall, discs were more slender in females, especially around early adolescence.Conclusion.The spine grows predominantly in the vertebral bodies. Thoracic discs increase in height only during the first years, whereas the transverse surface area continues to increase throughout growth, thus discs slenderness decreases. Relatively, female discs remained slenderer around growth-spurt. These measurements may assist future studies on the role of disc morphology in the etiology and treatment of spinal deformity.Level of Evidence: 4

    The diagnostic accuracy of F-18-FDG PET/CT in diagnosing fracture-related infections

    Get PDF
    PURPOSE: 18F-Fluorodeoxyglucose positron emission tomography (18F-FDG PET/CT) is frequently used to diagnose fracture-related infections (FRIs), but its diagnostic performance in this field is still unknown. The aims of this study were: (1) to assess the diagnostic performance of qualitative assessment of 18F-FDG PET/CT scans in diagnosing FRI, (2) to establish the diagnostic performance of standardized uptake values (SUVs) extracted from 18F-FDG PET/CT scans and to determine their associated optimal cut-off values, and (3) to identify variables that predict a false-positive (FP) or false-negative (FN) 18F-FDG PET/CT result. METHODS: This retrospective cohort study included all patients with suspected FRI undergoing 18F-FDG PET/CT between 2011 and 2017 in two level-1 trauma centres. Two nuclear medicine physicians independently reassessed all 18F-FDG PET/CT scans. The reference standard consisted of the result of at least two deep, representative microbiological cultures or the presence/absence of clinical confirmatory signs of FRI (AO/EBJIS consensus definition) during a follow-up of at least 6 months. Diagnostic performance in terms of sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) was calculated. Additionally, SUVs were measured on 18F-FDG PET/CT scans. Volumes of interest were drawn around the suspected and corresponding contralateral areas to obtain absolute values and ratios between suspected and contralateral areas. A multivariable logistic regression analysis was also performed to identify the most important predictor(s) of FP or FN 18F-FDG PET/CT results. RESULTS: The study included 156 18F-FDG PET/CT scans in 135 patients. Qualitative assessment of 18F-FDG PET/CT scans showed a sensitivity of 0.89, specificity of 0.80, PPV of 0.74, NPV of 0.91 and diagnostic accuracy of 0.83. SUVs on their own resulted in lower diagnostic performance, but combining them with qualitative assessments yielded an AUC of 0.89 compared to an AUC of 0.84 when considering only the qualitative assessment results (p = 0.007). 18F-FDG PET/CT performed <1 month after surgery was found to be the independent variable with the highest predictive value for a false test result, with an absolute risk of 46% (95% CI 27-66%), compared with 7% (95% CI 4-12%) in patients with 18F-FDG PET/CT performed 1-6 months after surgery. CONCLUSION: Qualitative assessment of 18F-FDG PET/CT scans had a diagnostic accuracy of 0.83 and an excellent NPV of 0.91 in diagnosing FRI. Adding SUV measurements to qualitative assessment provided additional accuracy in comparison to qualitative assessment alone. An interval between surgery and 18F-FDG PET/CT of <1 month was associated with a sharp increase in false test results
    corecore