    Prone vs. supine accelerated partial breast irradiation on an MR-Linac: A planning study

    Get PDF
    Background and purpose: Accelerated partial breast irradiation (APBI) may benefit from the MR-Linac for target definition, patient setup, and motion monitoring. In this planning study, we investigated whether prone or supine position is dosimetrically beneficial for APBI on an MR-Linac and we evaluated patient comfort. Materials and methods: Twenty-patients (9 postoperative, 11 preoperative) with a DCIS or breast tumor <3 cm underwent 1.5 T MRI in prone and supine position. The tumor or tumor bed was delineated as GTV and a 2 cm CTV-margin and 0.5 cm PTV-margin were added. 1.5 T MR-Linac treatment plans (5 Ă— 5.2 Gy) with 11 beams were created for both positions in each patient. We evaluated the number of plans that achieved the planning constraints and performed a dosimetric comparison between prone and supine position using the Wilcoxon signed-rank test (p-value <0.01 for significance). Patient experience during scanning was evaluated with a questionnaire. Results: All 40 plans met the target coverage and OAR constraints, regardless of position. Heart Dmean was not significantly different (1.07 vs. 0.79 Gy, p-value: 0.027). V5Gy to the ipsilateral lung (4.4% vs. 9.8% median, p-value 0.009) and estimated delivery time (362 vs. 392 s, p-value: 0.003) were significantly lower for prone position. PTV coverage and dose to other OAR were comparable between positions. The majority of patients (13/20) preferred supine position. Conclusion: APBI on the MR-Linac is dosimetrically feasible in prone and supine position. Mean heart dose was similar in both positions. Ipsilateral lung V5Gy was lower in prone position