Characterization of a Human Powered Nebulizer Compressor for Resource Poor Settings
Background
Respiratory disease accounts for three of the ten leading causes of death worldwide. Many of these diseases can be treated and diagnosed using a nebulizer. Nebulizers can also be used to safely and efficiently deliver vaccines. Unfortunately, commercially available nebulizers are not designed for use in regions of the world where lung disease is most prevalent: they are electricity-dependent, cost-prohibitive, and not built to be reliable in harsh operating conditions or under frequent use.
To overcome these limitations, the Human Powered Nebulizer compressor (HPN) was developed. The HPN does not require electricity; instead airflow is generated manually through a hand-crank or bicycle-style pedal system. A health care worker or other trained individual operates the device while the patient receives treatment.
This study demonstrates functional specifications of the HPN in comparison with a standard commercially available electric jet nebulizer compressor, the DeVilbiss Pulmo-Aide 5650D (Pulmo-Aide). Methods
Pressure and flow characteristics were measured with a rotameter and pressure transducer, respectively. Volume nebulized by each compressor was determined by mass, and particle size distribution was determined via laser diffraction. The Hudson RCI Micro Mist nebulizer mouthpiece was used with both compressors. Results
The pressure and flow generated by the HPN and Pulmo-Aide were: 15.17 psi and 10.5 L/min; and 14.65 psi and 11.2 L/min, respectively. The volume of liquid delivered by each was equivalent, 1.097 ± 0.107 mL (mean ± s.e.m., n = 13) for the HPN and 1.092 ± 0.116 mL for the Pulmo-Aide. The average particle size was also equivalent, 5.38 ± 0.040 micrometers (mean ± s.e.m., n = 7) and 5.40 ± 0.025 micrometers, respectively. Conclusions
Based on these characteristics, the HPN’s performance is equivalent to a popular commercially available electric nebulizer compressor. The findings presented in this paper, combined with the results of two published clinical studies, suggest that the HPN could serve as an important diagnostic and therapeutic tool in the fight against global respiratory health challenges including: tuberculosis, chronic obstructive pulmonary disease, asthma, and lower respiratory infections
Hyperglycemia Is Associated With Enhanced Thrombin Formation, Platelet Activation, and Fibrin Clot Resistance to Lysis in Patients With Acute Coronary Syndrome
OBJECTIVE—Acute hyperglycemia on admission for acute coronary syndrome worsens the prognosis in patients with and without known diabetes. Postulated mechanisms of this observation include prothrombotic effects. The aim of this study was to evaluate the effect of elevated glucose levels on blood clotting in acute coronary syndrome patients
Cardiopoietic cell therapy for advanced ischemic heart failure: results at 39 weeks of the prospective, randomized, double blind, sham-controlled CHART-1 clinical trial
Cardiopoietic cells, produced through cardiogenic conditioning of patients' mesenchymal stem cells, have shown preliminary efficacy. The Congestive Heart Failure Cardiopoietic Regenerative Therapy (CHART-1) trial aimed to validate cardiopoiesis-based biotherapy in a larger heart failure cohort