    The ARMM System-Autonomous Steering of Magnetically-Actuated Catheters:Towards Endovascular Applications

    Get PDF
    Positioning conventional endovascular catheters is not without risk, and there is a multitude of complications that are associated with their use in manual surgical interventions. By utilizing surgical manipulators, the efficacy of remote-controlled catheters can be investigated in vivo. However, technical challenges, such as the duration of catheterizations, accurate positioning at target sites, and consistent imaging of these catheters using non-hazardous modalities, still exist. In this paper, we propose the integration of multiple sub-systems in order to extend the clinical feasibility of an autonomous surgical system designed to address these challenges. The system handles the full synchronization of co-operating manipulators that both actuate a clinical tool. The experiments within this study are conducted within a clinically-relevant workspace and inside a gelatinous phantom that represents a life-size human torso. A catheter is positioned using magnetic actuation and proportional-integral (PI) control in conjunction with real-time ultrasound images. Our results indicate an average error between the tracked catheter tip and target positions of 2:09 0:49 mm. The median procedure time to reach targets is 32:6 s. We expect that our system will provide a step towards collaborative manipulators employing mobile electromagnets, and possibly improve autonomous catheterization procedures within endovascular surgeries