4 research outputs found

    Comparison of Jugular vs. Saphenous Blood Samples, Intrarater and In-Between Device Reliability of Clinically Used ROTEM S Parameters in Dogs

    Full text link
    Rotational Thromboelastometry (ROTEM) allows for the global assessment of hemostasis in whole blood samples. Preanalytical and analytical factors may influence test results, and data about the reliability and reproducibility of lyophilized ROTEM tests are scarce. Therefore, the objective of this study was to evaluate the influence of blood collection site on ROTEM S parameters and to assess intrarater and in-between device variability. A total of thirty, healthy, staff-owned dogs were included. Blood collection and ROTEM analysis were performed by trained staff according to a standardized protocol. Extrinsically activated (tissue factor; Ex-TEM S), with the addition of cytochalasin for platelet inhibition (Fib-TEM S), and intrinsically activated (In-TEM) analyses were performed. Analysis of our data showed significant variability for various Ex-TEM S and Fib-TEM S parameters from different collection sites and intrarater and in-between device measurements. We conclude that serial monitoring with ROTEM should be performed on the same device, with blood always taken from the same collection site using a standardized blood sampling technique. While In-TEM S, apart from maximum lysis, showed very stable and reliable results, we suggest interpreting especially clotting and clot formation parameters from Ex-TEM S and Fib-TEM S tests with caution and using duplicate measurements to detect outliers and to prevent initiation of incorrect therapies

    Evaluation of the Effect of Storage Time on ROTEM S® Parameters in Healthy and Ill Dogs

    Full text link
    Viscoelastic testing as a bedside test to assess global haemostasis has gained popularity in the past decade, with rotational thromboelastometry (ROTEM) and thromboelastography (TEG) being the two commonly used devices. TEG studies suggest analysis 30 min after blood sampling. However, the reproducibility of results over time for ROTEM analysis using lyophilized samples in dogs has not been established. In this study, we investigated the influence of time on viscoelastic testing, using 33 healthy staff-/client-owned dogs for blood sampling and repeated measurements of ROTEM tracings at three different time points after blood collection. Additionally, a group of 21 hospitalized patients with suspected coagulation disorders were included to investigate whether stability over time was comparable between healthy and ill dogs. We demonstrated a significant difference of ROTEM tracings over time, with a tendency towards hypocoagulability over time. These changes do have a clinical relevance as they exceed reference intervals and could therefore lead to erroneous conclusions about a patient’s coagulation status. Therefore, time-specific reference intervals are proposed and presented in this publication

    Evaluation of the Effect of Storage Time on ROTEM S® Parameters in Healthy and Ill Dogs.

    Get PDF
    Viscoelastic testing as a bedside test to assess global haemostasis has gained popularity in the past decade, with rotational thromboelastometry (ROTEM) and thromboelastography (TEG) being the two commonly used devices. TEG studies suggest analysis 30 min after blood sampling. However, the reproducibility of results over time for ROTEM analysis using lyophilized samples in dogs has not been established. In this study, we investigated the influence of time on viscoelastic testing, using 33 healthy staff-/client-owned dogs for blood sampling and repeated measurements of ROTEM tracings at three different time points after blood collection. Additionally, a group of 21 hospitalized patients with suspected coagulation disorders were included to investigate whether stability over time was comparable between healthy and ill dogs. We demonstrated a significant difference of ROTEM tracings over time, with a tendency towards hypocoagulability over time. These changes do have a clinical relevance as they exceed reference intervals and could therefore lead to erroneous conclusions about a patient's coagulation status. Therefore, time-specific reference intervals are proposed and presented in this publication

    Comparison of Jugular vs. Saphenous Blood Samples, Intrarater and In-Between Device Reliability of Clinically Used ROTEM S Parameters in Dogs.

    Get PDF
    Rotational Thromboelastometry (ROTEM) allows for the global assessment of hemostasis in whole blood samples. Preanalytical and analytical factors may influence test results, and data about the reliability and reproducibility of lyophilized ROTEM tests are scarce. Therefore, the objective of this study was to evaluate the influence of blood collection site on ROTEM S parameters and to assess intrarater and in-between device variability. A total of thirty, healthy, staff-owned dogs were included. Blood collection and ROTEM analysis were performed by trained staff according to a standardized protocol. Extrinsically activated (tissue factor; Ex-TEM S), with the addition of cytochalasin for platelet inhibition (Fib-TEM S), and intrinsically activated (In-TEM) analyses were performed. Analysis of our data showed significant variability for various Ex-TEM S and Fib-TEM S parameters from different collection sites and intrarater and in-between device measurements. We conclude that serial monitoring with ROTEM should be performed on the same device, with blood always taken from the same collection site using a standardized blood sampling technique. While In-TEM S, apart from maximum lysis, showed very stable and reliable results, we suggest interpreting especially clotting and clot formation parameters from Ex-TEM S and Fib-TEM S tests with caution and using duplicate measurements to detect outliers and to prevent initiation of incorrect therapies