    Optimal measurement counting time and statistics in gamma spectrometry analysis: The time balance

    Full text link
    The optimal measurement counting time for gamma-ray spectrometry analysis using HPGe detectors was determined in our laboratory by comparing twelve hours measurement counting time at day and twelve hours measurement counting time at night. The day spectrum does not fully cover the night spectrum for the same sample. It is observed that the perturbation come to the sun-light. After several investigations became clearer: to remove all effects of radiation from outside (earth, the sun, and universe) our system, it is necessary to measure the background for 24, 48 or 72 hours. In the same way, the samples have to be measured for 24, 48 or 72 hours to be safe to be purified the measurement (equality of day and night measurement). It is also possible to not use the background of the winter in summer. Depend on to the energy of radionuclide we seek, it is clear that the most important steps of a gamma spectrometry measurement are the preparation of the sample and the calibration of the detector