    胸壁外に設置した拍動型補助人工心臓に関する前臨床的研究

    Get PDF
    An assistance artificial heart was fabricated with blood compatible polyurethane having 3 different sizes and was evaluated by a series of preclinical experiment in calves. The following results were obtained. 1) The artificial heart was constructed in 3 sizes of 60, 80 and 100 cc in capacity and is activated by a pneumatic heart driver console. 2) The mock circulation revealed that this assistance heart has ejection fraction of 80-95%. 3) Safe use of assistance heart has been confirmed by implantation in 18 calves with intact or failured hearts over a period of 3 months. 4) Weaning could be performed in cases where myocardial necrosis of the left ventricle is induced less than 35% utilizing myocardial injection of 5N-NaOH. However pump dependency develops for a necrosis exceeding 35 %. 5) Thus we concluded that this system can be safely employed for a one month cardiac assistance clinically and at most for a 3 months support in pump dependent cases. 6) The technique was developed to make the instant shift possible within a few minutes from the conventional non-pulsatile bypass assistance to the assistance heart