Simultaneous Online Monitoring of Multiple Reactions Using a Miniature Mass Spectrometer
Advances in chemical
sampling using miniature mass spectrometer
technology are used to monitor slow reactions at a frequency of ca.
180 h<sup>–1</sup> (on the Mini 12) with no sample carryover
and with inline derivatization in the case of poorly ionizing compounds.
Moreover, we demonstrate high reproducibility with a relative error
of less than 10% for major components. Monitoring is enabled using
a continuous-flow nanoelectrospray (CF-nESI) probe contained in a
custom-built 3D-printed rotary holder. The holder position is automatically
set using a stepper motor controlled by a microcontroller. Reaction
progress of up to six reactions, including hydrazone formation and
Katritzky transamination, can be monitored simultaneously without
carryover for several hours