    Case report: High-resolution, intra-operative µDoppler-imaging of spinal cord hemangioblastoma

    Get PDF
    Surgical resection of spinal cord hemangioblastomas remains a challenging endeavor: the neurosurgeon’s aim to reach total tumor resections directly endangers their aim to minimize post-operative neurological deficits. The currently available tools to guide the neurosurgeon’s intra-operative decision-making consist mostly of pre-operative imaging techniques such as MRI or MRA, which cannot cater to intra-operative changes in field of view. For a while now, spinal cord surgeons have adopted ultrasound and its submodalities such as Doppler and CEUS as intra-operative techniques, given their many benefits such as real-time feedback, mobility and ease of use. However, for highly vascularized lesions such as hemangioblastomas, which contain up to capillary-level microvasculature, having access to higher-resolution intra-operative vascular imaging could potentially be highly beneficial. µDoppler-imaging is a new imaging modality especially fit for high-resolution hemodynamic imaging. Over the last decade, µDoppler-imaging has emerged as a high-resolution, contrast-free sonography-based technique which relies on High-Frame-Rate (HFR)-ultrasound and subsequent Doppler processing. In contrast to conventional millimeter-scale (Doppler) ultrasound, the µDoppler technique has a higher sensitivity to detect slow flow in the entire field-of-view which allows for unprecedented visualization of blood flow down to sub-millimeter resolution. In contrast to CEUS, µDoppler is able to image high-resolution details continuously, without being contrast bolus-dependent. Previously, our team has demonstrated the use of this technique in the context of functional brain mapping during awake brain tumor resections and surgical resections of cerebral arteriovenous malformations (AVM). However, the application of µDoppler-imaging in the context of the spinal cord has remained restricted to a handful of mostly pre-clinical animal studies. Here we describe the first application of µDoppler-imaging in the case of a patient with two thoracic spinal hemangioblastomas. We demonstrate how µDoppler is able to identify intra-operatively and with high-resolution, hemodynamic features of the lesion. In contrast to pre-operative MRA, µDoppler could identify intralesional vascular details, in real-time during the surgical procedure. Additionally, we show highly detailed post-resection images of physiological human spinal cord anatomy. Finally, we discuss the necessary future steps to push µDoppler to reach actual clinical maturity

    The Dorsal Root Ganglion as a Novel Neuromodulatory Target to Evoke Strong and Reproducible Motor Responses in Chronic Motor Complete Spinal Cord Injury: A Case Series of Five Patients

    Get PDF
    Objectives: Current strategies for motor recovery after spinal cord injury (SCI) aim to facilitate motor performance through modulation of afferent input to the spinal cord using epidural electrical stimulation (EES). The dorsal root ganglion (DRG) itself, the first relay station of these afferent inputs, has not yet been targeted for this purpose. The current study aimed to determine whether DRG stimulation can facilitate clinically relevant motor response in motor complete SCI. Materials and Methods: Five patients with chronic motor complete SCI were implanted with DRG leads placed bilaterally on level L4 during five days. Based on personalized stimulation protocols, we aimed to evoke dynamic (phase 1) and isotonic (phase 2) motor responses in the bilateral

    Functional Ultrasound (fUS) During Awake Brain Surgery: The Clinical Potential of Intra-Operative Functional and Vascular Brain Mapping

    Get PDF
    Background and Purpose: Oncological neurosurgery relies heavily on making continuous, intra-operative tumor-brain delineations based on image-guidance. Limitations of currently available imaging techniques call for the development of real-time image-guided resection tools, which allow for reliable functional and anatomical information in an intra-operative setting. Functional ultrasound (fUS), is a new mobile neuro-imaging tool with unprecedented spatiotemporal resolution, which allows for the detection of small changes in blood dynamics that reflect changes in metabolic activity of activated neurons through neurovascular coupling. We have applied fUS during conventional awake brain surgery to determine its clinical potential for both intra-operative functional and vascular brain mapping, with the ultimate aim of achieving maximum safe tumor resection. Methods: During awake brain surgery, fUS was used to image tumor vasculature and task-evoked brain activation with electrocortical stimulation mapping (ESM) as a gold standard. For functional imaging, patients were presented with motor, language or visual tasks, while the probe was placed over (ESM-defined) functional brain areas. For tumor vascular imaging, tumor tissue (pre-resection) and tumor resection cavity (post-resection) were imaged by moving the hand-held probe along a continuous trajectory over the regions of interest. Results: A total of 10 patients were included, with predominantly intra-parenchymal frontal and temporal lobe tumors of both low and higher histopathological grades. fUS was able to detect (ESM-defined) functional areas deep inside the brain for a range of functional tasks including language processing. Brain tissue could be imaged at a spatial and temporal resolution of 300 μm and 1.5–2.0 ms respectively, revealing real-time tumor-specific, and healthy vascular characteristics. Conclusion: The current study presents the potential of applying fUS during awake brain surgery. We i

    Dorsal root ganglion (DRG) versatile stimulator prototype developed for use in locomotion recovery early clinical trials

    No full text
    This paper presents the development of a Dorsal Root Ganglion (DRG) stimulator system intended for use in early clinical trials for motor recovery after Spinal Cord Injury (SCI). It allows for independent control of multisite/multilevel bilateral (on both sides of the spinal cord) stimulation, it can supply a high output current of 25.4mA, and has the ability to program pulse sequences similar to actual locomotion patterns. These characteristics ultimately provide the required versatility for examining the effects of DRG stimulation on locomotion recovery, which is lacking in currently available commercial systems. The device is created using commercially available components to make the design reproducible by other research labs and to facilitate the critical approval procedure for use in a clinical research environment. Throughout the design phase, essential considerations regarding the safety of the participating patient, as well as of the medical personnel involved, were taken into account and these are analyzed and demonstrated in this paper. Such considerations are very rarely discussed in scientific literature and the authors consider that, apart from the design of the system itself, this discussion is a critical contribution of this paper. Accepted author manuscriptBio-Electronic

    The Dorsal Root Ganglion as a Novel Neuromodulatory Target to Evoke Strong and Reproducible Motor Responses in Chronic Motor Complete Spinal Cord Injury: A Case Series of Five Patients

    No full text
    OBJECTIVES: Current strategies for motor recovery after spinal cord injury (SCI) aim to facilitate motor performance through modulation of afferent input to the spinal cord using epidural electrical stimulation (EES). The dorsal root ganglion (DRG) itself, the first relay station of these afferent inputs, has not yet been targeted for this purpose. The current study aimed to determine whether DRG stimulation can facilitate clinically relevant motor response in motor complete SCI. MATERIALS AND METHODS: Five patients with chronic motor complete SCI were implanted with DRG leads placed bilaterally on level L4 during five days. Based on personalized stimulation protocols, we aimed to evoke dynamic (phase 1) and isotonic (phase 2) motor responses in the bilateral quadriceps muscles. On days 1 and 5, EMG-measurements (root mean square [RMS] values) and clinical muscle force measurements (MRC scoring) were used to measure motor responses and their reproducibility. RESULTS: In all patients, DRG-stimulation evoked significant phase 1 and phase 2 motor responses with an MRC ≥4 for all upper leg muscles (rectus femoris, vastus lateralis, vastus medialis, and biceps femoris) (p < 0.05 and p < 0.01, respectively), leading to a knee extension movement strong enough to facilitate assisted weight bearing. No significant differences in RMS values were observed between days 1 and 5 of the study, indicating that motor responses were reproducible. CONCLUSION: The current paper provides first evidence that bilateral L4 DRG stimulation can evoke reproducible motor responses in the upper leg, sufficient for assisted weight bearing in patients with chronic motor complete SCI. As such, a new target for SCI treatment has surfaced, using existing stimulation devices, making the technique directly clinically accessible

    Case report: High-resolution, intra-operative µDoppler-imaging of spinal cord hemangioblastoma

    Get PDF
    Surgical resection of spinal cord hemangioblastomas remains a challenging endeavor: the neurosurgeon’s aim to reach total tumor resections directly endangers their aim to minimize post-operative neurological deficits. The currently available tools to guide the neurosurgeon’s intra-operative decision-making consist mostly of pre-operative imaging techniques such as MRI or MRA, which cannot cater to intra-operative changes in field of view. For a while now, spinal cord surgeons have adopted ultrasound and its submodalities such as Doppler and CEUS as intra-operative techniques, given their many benefits such as real-time feedback, mobility and ease of use. However, for highly vascularized lesions such as hemangioblastomas, which contain up to capillary-level microvasculature, having access to higher-resolution intra-operative vascular imaging could potentially be highly beneficial. µDoppler-imaging is a new imaging modality especially fit for high-resolution hemodynamic imaging. Over the last decade, µDoppler-imaging has emerged as a high-resolution, contrast-free sonography-based technique which relies on High-Frame-Rate (HFR)-ultrasound and subsequent Doppler processing. In contrast to conventional millimeter-scale (Doppler) ultrasound, the µDoppler technique has a higher sensitivity to detect slow flow in the entire field-of-view which allows for unprecedented visualization of blood flow down to sub-millimeter resolution. In contrast to CEUS, µDoppler is able to image high-resolution details continuously, without being contrast bolus-dependent. Previously, our team has demonstrated the use of this technique in the context of functional brain mapping during awake brain tumor resections and surgical resections of cerebral arteriovenous malformations (AVM). However, the application of µDoppler-imaging in the context of the spinal cord has remained restricted to a handful of mostly pre-clinical animal studies. Here we describe the first application of µDoppler-imaging in the case of a patient with two thoracic spinal hemangioblastomas. We demonstrate how µDoppler is able to identify intra-operatively and with high-resolution, hemodynamic features of the lesion. In contrast to pre-operative MRA, µDoppler could identify intralesional vascular details, in real-time during the surgical procedure. Additionally, we show highly detailed post-resection images of physiological human spinal cord anatomy. Finally, we discuss the necessary future steps to push µDoppler to reach actual clinical maturity

    The Dorsal Root Ganglion as a Novel Neuromodulatory Target to Evoke Strong and Reproducible Motor Responses in Chronic Motor Complete Spinal Cord Injury: A Case Series of Five Patients

    Get PDF
    OBJECTIVES: Current strategies for motor recovery after spinal cord injury (SCI) aim to facilitate motor performance through modulation of afferent input to the spinal cord using epidural electrical stimulation (EES). The dorsal root ganglion (DRG) itself, the first relay station of these afferent inputs, has not yet been targeted for this purpose. The current study aimed to determine whether DRG stimulation can facilitate clinically relevant motor response in motor complete SCI. MATERIALS AND METHODS: Five patients with chronic motor complete SCI were implanted with DRG leads placed bilaterally on level L4 during five days. Based on personalized stimulation protocols, we aimed to evoke dynamic (phase 1) and isotonic (phase 2) motor responses in the bilateral quadriceps muscles. On days 1 and 5, EMG-measurements (root mean square [RMS] values) and clinical muscle force measurements (MRC scoring) were used to measure motor responses and their reproducibility. RESULTS: In all patients, DRG-stimulation evoked significant phase 1 and phase 2 motor responses with an MRC ≥4 for all upper leg muscles (rectus femoris, vastus lateralis, vastus medialis, and biceps femoris) (p < 0.05 and p < 0.01, respectively), leading to a knee extension movement strong enough to facilitate assisted weight bearing. No significant differences in RMS values were observed between days 1 and 5 of the study, indicating that motor responses were reproducible. CONCLUSION: The current paper provides first evidence that bilateral L4 DRG stimulation can evoke reproducible motor responses in the upper leg, sufficient for assisted weight bearing in patients with chronic motor complete SCI. As such, a new target for SCI treatment has surfaced, using existing stimulation devices, making the technique directly clinically accessible
    corecore