Beam tests of the gas electron multiplier
We describe the results of systematic measurements, carried out with single and double GEM detectors with printed circuit read-out and having an active area 10x10 cm , both in the laboratory and in a high energy charged particles beam at CERN. Using fast analogue readout electronics, we demonstrate efficiencies for minimum ionizing particles close to 100%, with typical signal/noise ratios above 50 and up to 10 for the single and double GEM configuration, respectively, and a time resolution of 15 ns fwhm. Localization accuracies around 40 mm rms have been obtained for perpendicular tracks, degrading to 200 mm at 20° of incidence to the normal. Operated in a non-flammable gas mixture (argon-carbon dioxide), GEM detectors are robust, light and cheap to manufacture, and offer excellent performances and reliability suited for use in the harsh environments met at high luminosity colliders
Effects of High Charge Densities in Multi-GEM Detectors
A comprehensive study, supported by systematic measurements and numerical
computations, of the intrinsic limits of multi-GEM detectors when exposed to
very high particle fluxes or operated at very large gains is presented. The
observed variations of the gain, of the ion back-flow, and of the pulse height
spectra are explained in terms of the effects of the spatial distribution of
positive ions and their movement throughout the amplification structure. The
intrinsic dynamic character of the processes involved imposes the use of a
non-standard simulation tool for the interpretation of the measurements.
Computations done with a Finite Element Analysis software reproduce the
observed behaviour of the detector. The impact of this detailed description of
the detector in extreme conditions is multiple: it clarifies some detector
behaviours already observed, it helps in defining intrinsic limits of the GEM
technology, and it suggests ways to extend them.Comment: 5 pages, 6 figures, 2015 IEEE Nuclear Science Symposiu
Aging measurements with the gas electron multiplier (GEM)
Continuing previous aging measurements with detectors based on the Gas Electron Multiplier (GEM), a cm triple-GEM detector, as used in the small area tracking of the COMPASS experiment at CERN, was investigated. With a detector identical to those installed in the experiment, long-term, high-rate exposures to keV X-ray radiation were performed to study its aging properties. In standard operation conditions, with Ar:CO (70:30) filling and operated at an effective gain of , no change in gain and energy resolution is observed after collecting a total charge of 7mC/mm, corresponding to seven years of normal operation. This observation confirms previous results demonstrating the relative insensitivity of GEM detectors to aging, even when manufactured with common materials
Optimisation of the Gas Electron Multiplier for high rate application
The construction and performance of large size GEM~detectors for the COMPASS~experiment is described. Based on the experience gained during the operation of these detectors in high rate muon, proton, and pion beams we discuss the suitability of their use in harsh radiation environments
Charge Transfer Properties Through Graphene for Applications in Gaseous Detectors
Graphene is a single layer of carbon atoms arranged in a honeycomb lattice
with remarkable mechanical and electrical properties. Regarded as the thinnest
and narrowest conductive mesh, it has drastically different transmission
behaviours when bombarded with electrons and ions in vacuum. This property, if
confirmed in gas, may be a definitive solution for the ion back-flow problem in
gaseous detectors. In order to ascertain this aspect, graphene layers of
dimensions of about 2x2cm, grown on a copper substrate, are transferred
onto a flat metal surface with holes, so that the graphene layer is freely
suspended. The graphene and the support are installed into a gaseous detector
equipped with a triple Gaseous Electron Multiplier (GEM), and the transparency
properties to electrons and ions are studied in gas as a function of the
electric fields. The techniques to produce the graphene samples are described,
and we report on preliminary tests of graphene-coated GEMs.Comment: 4pages, 3figures, 13th Pisa Meeting on Advanced Detector
New observations with the gas electron multiplier (GEM)
We describe recent measurements realized with the Gas Electron Multiplier (GEM) mesh added as pre-amplification element to a multiwire and a micro-strip chamber. Large, stable combined gains are obtained, with good uniformity and energy resolution, in a wide range of filling gases including non-flammable mixtures; coupled to a micro-strip plate, the pre-amplification element allows to maintain the high rate capability and resolution at considerably lower operating voltages, completely eliminating discharge problems. Charge gains are large enough to allow detection of signals in the ionization mode on the last element, permitting the use of a simple printed circuit as read-out electrode; two-dimensional read-out can then be easily implemented. The absence of charge multiplication in the last stage avoids charge build-up on the substrate and prevents ageing phenomena. A new generation of simple, reliable and cheap fast position sensitive detectors seems at hand
The gas electron multiplier (GEM)
We describe operating priciples and results obtained with a new detector component: the Gas Electrons Multiplier (GEM). Consisting of a thin composite sheet with two metal layers separated by a thin insulator, and pierced by a regular matrix of open channels, the GEM electrode, inserted on the path of electrons in a gas detector, allows to transfer the charge with an amplification factor approaching ten. Uniform response and high rate capability are demonstrated. Coupled to another device, multiwire or micro-strip chamber, the GEM electrode permit to obtain higher gains or less critical operation; separation of the sensitive (conversion) volume and the detection volume has other advantages, as a built-in delay (useful for triggering purposes) and the possibility of applying high fields on the photo-cathode of ring imaging detectors to improve efficiency. Multiple GEM grids in the same gas volume allow to obtain large amplification factors in a succession of steps, leading to the realization of an effective gas-filled photomultiplier
Further developments of the gas electron multiplier (GEM)
We describe the development and operation of the Gas Electron Multiplier, a thin insulating foil metal-clad on both sides and perforated by a regular pattern of small holes. The mesh can be incorporat ed in the gas volume of an active detector to provide a first amplification channel for electrons, or used as stand alone. We report on the basic properties of GEMs manufactured with different geometr ies and operated in several gas mixtures as well as on their long-term stability after accumulation of charge equivalent to several years of operation in high luminosity experiments. Optimized GEMs re ach gains close to 10000 at safe operating voltages, permitting the detection of ionizing tracks, without other amplifying elements, on a simple printed circuit board (PCB), opening new possibilities for detector design
Two-dimensional readout of GEM detectors
The recently introduced Gas Electron Multiplier (GEM) permits the amplification of electrons released by ionizing radiation in a gas by factors approaching ten thousand, larger gains can be obtained combining two GEMs in cascade. We describe methods for implementing two- and three-dimensional projective localization of radiation, with sub-millimeter accuracy, making use of specially manufactured and patterned pick-up electrodes. Easy to implement and flexible in the choice of the readout geometry, the technology has the distinctive advantage of allowing all pick-up electrodes to be kept at ground potential, thus substantially improving the system simplicity and reliability. Preliminary results demonstrating the two-dimensional imaging capability of the devices are provided and discussed, as well as future perspectives of development
Optimization of design and beam test of microstrip gas chambers
We describe recent experimental and theoretical work aimed at optimizing the geometry and the operation of micro-strip gas chambers in order to improve their performance and reliability. With the help of a simulation program, we have studied the mechanism of signal propagation and analyzed the effects on signal shape and size of resistivity of strips, grouping of biased strips and presence of a back-plane. Several detectors manufactured according to the results of the study and equipped with fast amplifiers have been installed in a test beam to study general operating characteristics, efficiency and localization accuracy; preliminary results of the data analysis are discussed
- …